Setting the Calibration and Adjustment

7.2.1
Switching the isoCAL Function On or Off
(Only Model i-1x)
When using the isoCAL function, the device performs an
automatic time- and temperature-dependent internal
calibration and adjustment .
If this relates to a conformity-assessed device in
legal metrology: In some cases it is not possible to
switch off the isoCAL function .
Procedure
Open the "SETUP"/"BALANCE" menu .
t
To set the automatic start of the isoCAL function: Select the
t
"ON" calibration value for the "ISOCAL" parameter .
To set the manual start of the isoCAL function: Select the
t
"Note" calibration value for the "ISOCAL" parameter .
To switch off the isoCAL function: Select the "OFF"
t
calibration value for the "ISOCAL" parameter .
7.2.2
Setting Internal Calibration and Adjustment
(Only Model i-1x)
The following functions can be set for the internal calibration
and adjustment:
− Internal calibration with automatic start of the adjustment .
− Internal calibration with manual start of the adjustment .
Procedure
Open the "SETUP"/"BALANCE" menu . Call up the "CAL .JUST ."
t
parameter and select the value "CAL .INT ." .
If the calibration function needs to be set with subsequent
t
automatic adjustment: In the "SETUP"/"BALANCE" menu, for
the "CAL .SEQ ." parameter, select the "ADJUST" setting value .
If the calibration function needs to be set without
t
subsequent automatic adjustment: In the
"SETUP"/"BALANCE" menu, for the "CAL .SEQ ." parameter,
select the "CAL ./ADJ ." setting value .
7.2.3
Setting the External Calibration and Adjustment
The following functions can be set for the external calibration
and adjustment:
− External calibration with manual start of the adjustment .
If this relates to a conformity-assessed device in
legal metrology: External calibration and adjust-
ment is not possible .
Procedure
Open the "SETUP"/"BALANCE" menu .
t
If the calibration function needs to be set with subsequent
t
automatic adjustment: Select the "ADJUST" calibration
value for the "CAL .SEQ ." parameter .
If the calibration function needs to be set without
t
subsequent automatic adjustment: Select the "CAL-ADJUST"
calibration value for the "CAL .SEQ ." parameter .
Setting the Weight Value for the External Weight
A preset weight value or a user-defined weight value can be set
for the external weight .
Procedure
If the preset weight value needs to be used: In the
t
"SETUP"/"BALANCE" menu, for the "CAL ./ADJ ." parameter,
select the "EXT .CAL ." setting value .
If a user-defined weight value needs to be set:
t
In the "INPUT" menu, select the "CAL .WT ." setting value .
t
Enter the desired weight value and press the [Confirm]
t
button .
To use the user-defined weight value for the next
t
calibration: In the "SETUP"/"BALANCE" menu, for the
"CAL ./ADJ ." parameter, select the "E .CAL .USR" setting
value .
Entris
®
System Settings
BCE Operating Instructions
17