Operating Concept
4.2.1

Buttons

Symbol
Designation
[On/Off] button
[Menu] button
[Zero] button
[Tare] button
[Print] button
[isoCAL] button
[Adjust] button
[GLP] button
[Ambient condition]
button
[Application filter]
button
[Toggle between
weight units] button
[Back] button
[Up] button
[Down] button
[Confirm] button
10
Entris
®
BCE Operating Instructions
Description
− When the button is pressed: Switches the operating display on .
− If the button is held down: Switches the operating display off .
− When the button is pressed: The settings menu opens .
− If the button is held down: It switches to version display .
Zeroes the device .
Starts taring .
Exports the readouts to the integrated data interfaces .
If the button is flashing: Starts the isoCAL function .
If the button is not flashing: Starts the set calibration and adjustment function .
Starts the set calibration and adjustment function .
− Exits the GLP printout and starts printing the GLP footer .
− If the "Net-total", "Totalizing", or "Statistics" application is active: Prints and deletes the
saved values and exits the application .
Switches between the ambient conditions "V .STABLE", "STABLE", UNSTABL ." and "V .UNSTBL ." .
Switches between the "weighing" and "filling" application filters .
If the "Toggle between weight units" function is active:
− If the button is held down: Accesses the "Toggle between weight units" function menu .
− When the button is pressed: Switches between the basic unit display and up to 4 other
units .
− In the menu:
− When the button is pressed: Returns to the previous display .
− If the button is held down: Saves the menu settings .
− When entering digits: Selects the previous digit position .
− For an active application: Exits the application and deletes the set reference value .
− In the menu: Scrolls through the menu levels or the available parameters .
− When entering digits: Increases the displayed value .
− In the main display of an active application: Switches to the display of the current weight
value/parameter .
− In the menu: Scrolls through the menu levels or the available parameters .
− When entering digits: Decreases the displayed value .
− In the main display of an application that is not active: Accesses the display to set the
reference values .
− In the main display of an active application: Switches to the display of the current weight
value/parameter .
− In the menu: Accesses the displayed menu level or confirms the displayed parameter .
− When entering digits: Selects the next digit position .
− In the main display of an application that is not active: Starts the application process and
saves the set reference value .
− In the main display of an active application: Adopts the next component or the next
parameter .

Quicklinks ausblenden: