Inspection intervals should be such that any significant
defects that may occur can be identified in good time.
Inspections should be carried out at least once every three
years. The applicable requirements and guidelines should also
be observed during operation.
Establish a low resistance connection from the scale and
the ex-link converter YCO16-Z to a customer-supplied
equipotential bonding conductor connection via the
equipotential bonding conductor connections (PA) on the
device using a suitable grounding cable with a gauge of at
2
least 4 mm
(not included).
Installation must be carried out properly by trained personnel
and according to commonly accepted technical standards.
The system should only be operated for the first time when
it is certain that the area is not potentially explosive.
If deviations are evident during startup due to transport
damage (e.g. no display, no backlighting), disconnect the scale
from the power supply and contact the Sartorius Service
Center.
Connect the scale to the equi-
potential bonding conductor using
an equipotential bonding cable
with a gauge of at least 4 mm
t
t
Connect the ex-link converter to
the equipotential bonding
conductor using another
equipotential bonding cable with
a gauge of at least 4 mm
t
t
Connecting a PC/Notebook
t
1
Connecting an AC Adapter
t
2
²
.
Connect the cable lug of the
equipotential bonding cable to
the grounding terminal of the
scale.
Connect the equipotential
bonding cable to the customer-
supplied equipotential bonding
conductor.
²
.
Connect the cable lug of the
equipotential bonding cable to
the grounding terminal of the
ex-link converter.
Connect the equipotential
bonding cable to the customer-
supplied equipotential bonding
conductor.

Plug an Ethernet cable (1)

into the Ex-Link converter
and connect the cable to
a Windows PC (direct
connection) or to the network.
Insert a USB cable (2) into the
ex-link converter.
t
t

3.5 Anti-theft Locking Device

t

3.6 Warm-up Time

0
To ensure accurate results are
delivered, the scale must warm up
for at least 30 minutes after initial
connection to the power supply.
Only after this time will the scale
have reached the required
operating temperature.
30
4 Accessing the Scale
Over the Network

4.1 Connecting to a Network with DHCP

Usually, the IP address on a network is assigned by a DHCP
server (Dynamic Host Configuration Protocol). A prerequisite
for this is that DHCP mode is enabled on the scale.
t
Press the "i" button to call up the "Settings" menu.
t
Go into the submenu "Ethernet".
If the display says "DHCP (On)", the settings are correct.
Installation Instructions VIS1X
English
Insert the USB cable into the
YPS07-USB AC adapter.
Plug the AC adapter into the
wall outlet (supply voltage).
If required, secure the scale at
the back.
7