System Settings

7
System Settings
7.1

Performing System Settings

Default settings can be adjusted for the device and the
applications in order to align with the ambient condi-
tions and individual operating requirements.
The following settings are necessary to operate the
device together with connected components:
− Set up the communication of the connected devices
− Set up additional components
The following settings are recommended to set up the
device:
− Set the menu language
− Set the date and time
− Set the calibration and adjustment
Procedure
Press the [Menu] button.
t
To adjust settings: Open the desired menu.
t
Select and confirm the desired parameter (parame-
t
ters, see Chapter "7.3 Parameter List", page 21).
Exit the menu.
t
7.2
Setting the Calibration and
Adjustment
7.2.1
Switching the isoCAL Function On or
Off (Only Model I-1x)
When using the isoCAL function, the device performs
an automatic time- and temperature-dependent inter-
nal calibration and adjustment.
If this relates to a conformity-assessed
device in legal metrology: In some cases it is
not possible to switch off the isoCAL
function.
20
Entris
BCE Operating Instructions
®
Procedure
Open the "SETUP" / "BALANCE" menu.
t
To set the automatic start of the isoCAL function:
t
Select the "ON" calibration value for the "ISOCAL"
parameter.
To set the manual start of the isoCAL function: Select
t
the "Note" calibration value for the "ISOCAL" param-
eter.
To switch off the isoCAL function: Select the "OFF"
t
calibration value for the "ISOCAL" parameter.
7.2.2
Setting Internal Calibration and
Adjustment (Only Model I-1x)
The following functions can be set for the internal cali-
bration and adjustment:
− Internal calibration with automatic start of the adjust-
ment.
− Internal calibration with manual start of the adjust-
ment.
Procedure
Open the "SETUP" / "BALANCE" menu. Call up the
t
"CAL.JUST." parameter and select the value "CAL.
INT.".
If the calibration function needs to be set with subse-
t
quent automatic adjustment: In the "SETUP" / "BAL-
ANCE" menu, for the "CAL.SEQ." parameter, select
the "ADJUST" setting value.
If the calibration function needs to be set without
t
subsequent automatic adjustment: In the
"SETUP" / "BALANCE" menu, for the "CAL.SEQ."
parameter, select the "CAL./ADJ." setting value.
7.2.3
Setting the External Calibration and
Adjustment
The following functions can be set for the external cali-
bration and adjustment:
− External calibration with manual start of the adjust-
ment.
If this relates to a conformity-assessed
device in legal metrology: External calibra-
tion and adjustment is not possible.