System Settings

7
System Settings
7.1

Performing System Settings

Default settings can be adjusted for the device and the
applications in order to align with the ambient
conditions and individual operating requirements.
The following settings are necessary to operate the
device together with connected components:
— Set up the communication of the connected devices
— Set up additional components
The following settings are recommended to set up the
device:
— Set the menu language
— Set the date and time
— Set the calibration and adjustment
Procedure
Press the [Menu] button.
t
To adjust settings: Open the desired menu.
t
Select and confirm the desired parameter
t
(parameters, see Kapitel „7.3 Parameter List",
Seite 19).
Exit the menu.
t
7.2
Setting the Calibration and
Adjustment
7.2.1
Switching the isoCAL Function On or
Off (Only Model I-1x)
When using the isoCAL function, the device performs
an automatic time- and temperature-dependent
internal calibration and adjustment.
If this relates to a conformity-assessed de-
vice in legal metrology: In some cases it is
not possible to switch off the isoCAL
function.
18
Entris
BCE Operating Instructions
®
Procedure
Open the "SETUP" / "BALANCE" menu.
t
To set the automatic start of the isoCAL function:
t
Select the "ON" calibration value for the "ISOCAL"
parameter.
To set the manual start of the isoCAL function: Select
t
the "Note" calibration value for the "ISOCAL"
parameter.
To switch off the isoCAL function: Select the "OFF"
t
calibration value for the "ISOCAL" parameter.
7.2.2
Setting Internal Calibration and
Adjustment (Only Model I-1x)
The following functions can be set for the internal
calibration and adjustment:
— Internal calibration with automatic start of the
adjustment.
— Internal calibration with manual start of the
adjustment.
Procedure
Open the "SETUP" / "BALANCE" menu. Call up the
t
"CAL.JUST." parameter and select the value "CAL.
INT.".
If the calibration function needs to be set with
t
subsequent automatic adjustment: In the
"SETUP" / "BALANCE" menu, for the "CAL.SEQ."
parameter, select the "ADJUST" setting value.
If the calibration function needs to be set without
t
subsequent automatic adjustment: In the
"SETUP" / "BALANCE" menu, for the "CAL.SEQ."
parameter, select the "CAL./ADJ." setting value.
7.2.3
Setting the External Calibration and
Adjustment
The following functions can be set for the external
calibration and adjustment:
— External calibration with manual start of the
adjustment.
If this relates to a conformity-assessed de-
vice in legal metrology: External calibration
and adjustment is not possible.