Herunterladen Inhalt Inhalt Diese Seite drucken

Calibration/Span Adjustment - Sartorius M-prove Betriebsanleitung

Inhaltsverzeichnis

Werbung

Verfügbare Sprachen
  • DE

Verfügbare Sprachen

  • DEUTSCH, seite 23

Calibration/Span Adjustment

Calibration is recommended after initial installation and each time the balance is moved.
Features
Calibration/adjustment can be performed only when:
– there is no load on the balance,
– the balance is zeroed, and
– the internal signal is stable.
If these conditions are not met, an error message is displayed.
The weight required for calibration/adjustment is displayed.
Standard calibration weights for selected models: to remove, see instructions on page 4.
Press (F) to select a different weight value.
To cancel the procedure: press and hold the (Enter) key (> 2 sec.).
Example: Calibrate/adjust span of the balance (here: model AY5101)
Step
1. Switch on the balance
2. Zero the balance
3. Start calibration
The preset calibration weight is displayed
without the weight unit
(in this example, 5000 g)
4. To select a different calibration weight value
5. Confirm calibration weight value and
start calibration/span adjustment
After the zero point is stored, the required
calibration weight flashes on the display.
6. Place the required calibration
weight on the balance
The readout stops flashing if the weight is
applied within the defined time limit and tolerance.
If the weight value is accepted, the display stops
flashing and the stability symbol
on the display.
7. Remove the calibration weight
8. Calibration has been completed
16
Key (or instruction)
e
(Zero)
(Cal)
(F) repeatedly
(Enter)
appears
Display

Quicklinks ausblenden:

Werbung

Inhaltsverzeichnis
loading

Inhaltsverzeichnis