Herunterladen Inhalt Inhalt Diese Seite drucken

Coltene BioSonic UC150 Benutzerhandbuch Seite 7

Ultraschallreinigungsgerät
Vorschau ausblenden Andere Handbücher für BioSonic UC150:
Inhaltsverzeichnis

Werbung

Verfügbare Sprachen
  • DE

Verfügbare Sprachen

  • DEUTSCH, seite 26
• CONNECT THE DISPLAY. Attach the connector on the bottom of the display module to the
connector on the top right-hand corner of the unit. Press the display module down carefully
until it snaps into the right position.
• PLUG IN THE UNIT. Attach the BioSonic UC150 electrical cord at the lower back of the unit and
plug into a three-wire grounded AC outlet (in accordance with local regulations) (Figure 9*). It is
recommended that the unit be plugged into a Ground Fault Circuit Interrupter (GFCI) outlet.
Installation instructions for recessing unit:
• Tools Required:
- Electric Drill
- Drill bit to drill a hole large enough to accept your reciprocating saw blade
- Reciprocating saw and blade
NOTE: This Ultrasonic Cleaner must be installed by an experienced equipment manager or
licensed contractor. Be sure to follow local building codes.
NOTE: Before you begin please read and understand the safety instructions provided by the
manufacturer of the tools required for installation.
• Select the desired location for the cleaner. You will need at least 19 cm (7.5 inches) clear vertical
space below the top of the counter to accommodate the tank enclosure. The countertop must
be flat and level. Make sure that you are not selecting a location which will interfere with
drawers or vertical partitions between cabinets.
• Tape the enclosed template down to the counter top in the desired location. Make sure that the
"front of the unit" is closest to the user.
• Cut along the "cut line" for the counter top with a jigsaw or other reciprocating saw.
• Clean away dirt or sawdust around the opening and remove scrap wood.
• Attach the rubber gasket on the housing for better seating between bezel and counter top.
• Lower unit into opening. Be sure that the printed BioSonic logo on the housing faces the user.
(Figure 11*)
• The hose supplied with the BioSonic UC150 will allow you to dispose of the waste in accordance
with your local, state and/or federal regulations.
*see last page for figures
7

Werbung

Inhaltsverzeichnis
loading

Inhaltsverzeichnis