14. Technical Details

Environmental Conditions:
• Intended for indoor use.
• Maximum altitude 2000 m.
• Recommended Storage and transport temperature:
• Temperature during operation:
• Relative humidity:
• Main supply voltage fluctuations not to exceed ± 10% of the nominal voltage.
• IEC-60664-1 Pollution degree 2.
Electrical Information:
• Power Supply Input: 100 V – 240 V AC / 50 ~ 60 Hz
• Transducers: 2
• Operating Frequency: 48 kHz - 50.5 kHz
• Fuse Specifications: 2 Fuses FST, 5 × 20 mm, Slow Blow, 250 VAC, 5A, IEC 60127-2 Sheet 3
Max Power @ Input Voltage
Input Voltage Max Power without Heater* Max Power with Heater*
100 V
160 W
115 V
160 W
230 V
160 W
240 V
160 W
*average over a sweep period
• EMC Class A: Suitable for use in all establishments other than domestic and those directly
connected to a low-voltage power supply network which supplies buildings used for domestic
purposes.
• Class I electrical equipment per IEC 61140.
425 W
510 W
510 W
540 W
Max Current
4.3 A
4.4 A
2.2 A
2.3 A