Error
Error Name
13
Over temperature Tank
14
Tank temperature rising
15
Tank temperature rising
20
Overcurrent
21
Overcurrent
22
Overvoltage
23
Overvoltage
24
Power deviation
25
Power deviation
26
Power deviation
27
Power deviation
28 - 39
N/A
40
Piezo no connection
41
Only one piezo
connected
42
Piezo short-circuit
43
Piezo no power
44
Main switch failure
45
Tank Temperature Sensor
failure
46
Voltage Measure failure
47
Current Measure failure
48
Phase Measure failure
49
N/A
50-69
Cable connection error
16
Description
Tank solution temperature is
too high
Solution bath temperature is
increasing too quickly
Solution bath temperature is
increasing too slowly
Abnormal current level
Abnormal current level
Abnormal voltage level
Abnormal voltage level
Power output is lower than
expected
Power output is lower than
expected
Power output is higher than
expected
Power output is higher than
expected
N/A
Unexpected error with the
piezo driver
Unexpected error with the
piezo driver
Unexpected error with the
piezo driver
Unexpected error with the
piezo driver
Main relay switch error
Temperature sensor not
responding
Unexpected drive signal
measurement
Unexpected drive signal
measurement
Unexpected drive signal
measurement
N/A
Internal cable failure
Recommended User Action
If the unit has been operated for extended
periods of time (e.g. multiple hours without
intermission) the bath temperature may rise
beyond an acceptable level. Either change the
solution or wait for it to cool. If the problem
persists, contact customer service.
Ensure that the tank is filled with solution to
within the fill lines. If problem persists, contact
customer service.
Ensure that the tank is filled with solution to
within the fill lines. If problem persists, contact
customer service.
Contact customer service.
Contact customer service.
Contact customer service.
Contact customer service.
Verify that the input voltage to the unit is
within the specifications noted in the
instructions for use. Deviations in supply
voltage may trigger this error. If the problem
persists, contact customer service.
Verify that the input voltage to the unit is
within the specifications noted in the
instructions for use. Deviations in supply
voltage may trigger this error. If the problem
persists, contact customer service.
Verify that the input voltage to the unit is
within the specifications noted in the
instructions for use. Deviations in supply
voltage may trigger this error. If the problem
persists, contact customer service.
Verify that the input voltage to the unit is
within the specifications noted in the
instructions for use. Deviations in supply
voltage may trigger this error. If the problem
persists, contact customer service.
N/A
Contact customer service.
Contact customer service.
Contact customer service.
Contact customer service.
Unplug the unit and contact customer service.
Do no attempt to reactivate the unit.
Contact customer service.
Contact customer service.
Contact customer service.
Contact customer service.
N/A
Contact customer service.