Required materials: 710D4 torque wrench, 636K13 Loctite®
>
1) Screw in the screwed insert (e.g. pyramid adapter) completely
into the thread of the lamination anchor.
2) The
screwed
insert
um of 1/4 turn out of the thread in order to align the insert.
3) If a titanium screw is on hand: After trial fitting, replace the
steel cap screw with the titanium cap screw.
4) Secure the cap screw of the threaded clamping using Loctite®
and tighten using the torque wrench (10 Nm).
Connecting the pyramid adapter and pyramid receiver
The pyramid adapter is fixed with the set screws of the pyramid receiv
er.
Required materials: 710D4 torque wrench, 636K13 Loctite® 241
>
1) Trial fitting:
Screw in the set screws.
Use the torque wrench to tighten the set screws (10 Nm).
2) Definitive mounting:
Use Loctite® to secure the set screws.
Screw in the set screws.
Pre-tighten the set screws with the torque wrench (10 Nm) and
then tighten them (15 Nm).
3) Replace any set screws that are protruding or recessed too much
with suitable ones (see selection table).
Selection table for set screws (length and material)
12 mm
Steel, galvan
506G3=M8X
ised
12-V
Stainless steel
506G3=M8X
12-"Niro"
Titanium
506G5=M8X
12
may
be
unscrewed
by
14 mm
506G3=M8X
506G3=M8X
14
506G3=M8X
506G3=M8X
14-"Niro"
16-"Niro"
506G5=M8X
506G5=M8X
14
Alignment
The set screws in the pyramid receiver can be used to make static
adjustments during alignment, trial fittings and after the prosthesis is
finished.
a
maxim
Replacement and disassembly
The set position of the prosthetic component can be maintained dur
ing replacement or disassembly. In order to do this, unscrew the two
set screws that are screwed in the furthest and located next to each
other.
6 Cleaning
1) Clean the product with a damp, soft cloth.
2) Dry the product with a soft cloth.
3) Allow to air dry in order to remove residual moisture.
7 Maintenance
CAUTION
Failure to follow the maintenance instructions
Risk of injuries due to changes in or loss of functionality and dam
age to the product
► Observe the following maintenance instructions.
► The prosthetic components should be inspected after the first 30
days of use.
► Inspect the entire prosthesis for wear during normal consulta
16 mm
tions.
► Conduct annual safety inspections.
16
8 Disposal
This product may not be disposed of with regular domestic waste in
all jurisdictions. Disposal that is not in accordance with the regula
tions of the country where the product is used may have a detrimental
16
impact on health and the environment. Please observe the information
17