Error messages

Error display (Err)
0002
0003
0004
0005
0006
0007
Cause
The limit value for relay alternations is
exceeded.
Terminal temperature
is outside the range -10...80°C
Reference voltage
The measured value either exceeds 999 or
goes below -999 and is outside the 3-digit
display range.
Calibration constant
Configuration data
Value cannot be displayed (too high or too
low)
reserved
reserved
15 Error messages
Remedy
h
Increase the limit value for the relay alternations.
v
Chapter 7.11 „C 120 Limit value for relay alterna-
tions"
h
Use the Reset key (press >3sec) for acknowl-
edgement.
v
Chapter 7.12 „C 121 Counter reading for relay
alternations"
h
Check ambient temperatures
h
Use the Reset key (press >3sec) for acknowl-
edgement.
If the error cannot be remedied, send the device in.
-
A/D converter error
h
Use the Reset key (press >3sec) to acknowledge
If the error cannot be remedied, send the device in.
The device must be repaired by JUMO.
h
Send in the device.
v
Chapter 2.1 „Service addresses"
-
-
54