Commissioning
®
Available in ReadWin
2000, HART
Commuwin II with Matrix address
PV lower range
Input of 4 mA value.
value
Input: Limitation values see device function SENSOR TYPE 1/2.
32 °F (0 °C)
PV upper
Input of 20 mA value.
range value
Input: Limitation values see device function SENSOR TYPE 1/2.
212 °F (100 °C)
Analog output
Input of the standard (4 to 20 mA) or inverse (20 to 4 mA) current output signal.
Input:
• 4 to 20 mA
• 20 to 4 mA
Filter
Selection of the digital filter 1. order (filter time constant).
Input: 0 to 60 s
HART Output/
Preamble
Multidrop
Device
address
®
Available in ReadWin
2000, HART
Commuwin II with Matrix address
Drift alert
Definition of the behavior if measured values for sensors 1 and 2 deviate from one
mode
another.
Input:
• off
• Warning
• Alarm
Warning:
The 'Caution' icon becomes active on the display. A warning is transmitted via the
®
HART
protocol.
Alarm:
The 'Caution' icon becomes active on the display. The device switches to error
signal.
Drift mode
Drift. No input necessary if the drift alert mode = off.
Input:
• Larger
Alarm/warning if absolute amount for difference between sensor 1 - sensor 2
overshoots a defined limit value (see drift alert value). Larger is the standard
value for device versions < SW 01.03.00 in which the parameter is not available.
• Smaller
Alarm/warning if absolute amount for difference between sensor 1 - sensor 2
undershoots a defined limit value (see drift alert value).
78
Function group OUTPUT
®
handheld module DXR275/375 (Symbol
Input: Number of response preambles: 0 to 15
5
Input: HART address of the temperature transmitters:
0 to 15
If addresses > 0, the temperature transmitter
is in Multidrop mode and the analog output is set to 4 mA. Device
address is shown on the display in the Multidrop mode
Function group SAFETY SETTINGS
®
handheld module DXR275/375 (Symbol
7
®
),
ReadWin
2000
+
+
+
+
-
7
®
),
ReadWin
2000
+
≥ SW 01.03.00
TMT162
7
/FieldCare
CW II-
Matrix
+
V1H4
+
V1H5
+
V1H6
+
V1H7
+
HART
Server
7
/FieldCare
CW II-
Matrix
+
V2H0
≥ SW 01.03.00
-