TMT162
Endress+Hauser
3.3.2
Remote installation
Fig. 4:
Installation of the field transmitter using the mounting kit, see chapter 'Accessories' (dimensions in in; mm)
A, B
Mounting with combined wall/pipe mounting kit
C
Mounting with pipe mounting kit 2"/V4A
3.4
Installation check
After installing the device, always run the following final checks:
Device condition and specification
Is the device visibly undamaged (visual check)?
Does the device comply with the measurement point specifications, such as
ambient temperature, measurement range, etc.?
Make sure that the transmitter covers are tight. Both transmitter covers must be
fully engaged to meet explosion-proof requirements.
Installation
a0003586-ae
Hint
-
See chapter 10 "Technical data"
-
63