During a multiple area treatment, after the treatment for one area commences, a pop-up screen will appear to prompt the user to move to the
next area and resume treatment. If the area being treated, has the option of treating bilaterally, the user will be prompted once all of the one-
sided treatments are complete.
The protocols are based on a series of Continuous Wave (CW) or Pulsed Rate combinations. The "Time" section on the left side
of the screen will display the time for the specific Pulsed Rate combination. The total time remaining will be displayed at the bottom
of the screen.
By selecting Max TX, which turns the Max TX button red (Figure 25), all of the Pulsed Rate (if any) settings will be converted to
Continuous Wave (CW). This will result in reduced treatment time.
5.5.
Operation Mode
Pressing Operation on the Main Menu will take the user to the Operation Screen and put the system in Standby Mode.
In Standby mode, the software allows the operator to change the treatment time, power, pulse frequency, and to reset
the energy counter (Figure 26).
5.6.
Operation Screen – Ready
In Ready mode, the laser will emit when the finger/foot switch is activated. The purpose of Ready mode is to wait for the user to press the
finger/foot switch to start laser emission. The software has a 6 second delay (6 beeps) to warn the user of the transition from Standby to
Ready mode.
Also, the Empower Emission Indicator Ring will flash yellow during the transition from Standby to Ready. While in Ready, the light will
remain yellow to indicate the laser state at the hand piece.
Warning: NEVER leave this device unattended in Ready mode.
To start laser emission, press the finger switch. To stop emission, remove finger from the finger switch or press finger switch again.
Note: If the unit is set up with finger switch override on, each time the finger switch is pressed, it will act as an on/off switch.
RSR-000524-000 (1)
Large Ball
Large Cone
Large attachments will only
appear Red or Green based
on the protocol and power
selection.
Figure 25
Figure 26
User Manual – CTX
Small Ball
Small Cone
Page 15 of 25