Dry with filtered totally oil-free compressed air
and make sure the contra-angle is thoroughly
dry before bagging it for sterilization.
Do not immerse the contra-angle handpiece
completely in detergent or disinfectant baths or
ultrasonic baths!
Insert the previously cleaned and disinfected
parts into the sterilization pouches and apply
the steam sterilization cycle at 134 °C for
3 minutes. The cycle has been validated for
EnDrive - Cleaning, Disinfection and Sterilization | 31
1,5 minutes. Use autoclaves that comply with
EN 13060.
Nextmed does not accept responsibility if these
instructions are not followed or if an unvalidated
process is used.
Use only properly maintained equipment and
personal protective equipment (gloves, goggles)
according to national laws and regulations.