10 | EnDrive - Warnings, precautions, contra-indications, adverse reactions
that the EnDrive components are function-
ing properly.
EnDrive requires precautions regarding
electromagnetic compatibility and must be
installed and operated in accordance with
the EMC Directive. In particular, EnDrive
must not be used in the vicinity of fluores-
cent lamps, radio transmitters, remote con-
trols or RF communication devices even if
they comply with the requirements of
CISPR11.
Ensure that the patient cable or
handpiece cable does not restrict the
operator's mobility during the procedure.
Do not knot the patient cable or handpiece
cable.
Do not pull on the cables but grasp the con-
nectors to disconnect them. Ensure that
the cables are never under tension.
The use of the EnDrive next to another
device should be avoided as this can lead
to malfunctions. Ensure that EnDrive and
the other device, if any, are operating nor-
mally.
Do not expose EnDrive to ultrasonic or
infrasonic sources.
Avoid direct light on the screen to ensure
full readability.
During treatment
The user must use their own clinical judge-
ment for appropriate use.
Ensure that the file selected on the screen
is the one actually in use on the contra-an-
gle handpiece. It is recommended to adhere
to the torque and speed settings shown on
the screen by default, as recommended by
the file manufacturer.
Do not use reciprocating files with continu-
ously rotating file settings, and vice versa.
Switch the main unit off via the on/off but-
ton in case it does not respond to com-
mands.
Lubricate the contra-angle handpiece after
each sterilisation cycle.
Do not lubricate the handpiece, as this will
damage the motor.
Make sure to calibrate the contra-angle
handpiece after each sterilisation and lubri-
cation cycle using the appropriate menu
and not during patient treatment. The cali-
bration process rotates the motor over the
entire speed range, so make sure the file is
not mounted on the contra-angle during
calibration.
If stopping the motor via the button on the
handpiece does not respond, press the
symbol of the instrument in use on the
main unit.
For the apex locator function, ensure that
the contact point between the endodontic
instrument and forceps is kept dry and
clean to avoid inaccurate readings.
The apex locator function during motor
treatment only works with the supplied con-
tra-angle handpiece.
If the file clamp is not properly connected
to the endodontic instrument, or is dam-
aged, EnDrive may generate incorrect infor-
mation.