Configuring The Probe Length; Programming Static Interference Signals - Sick Lfp Inox Betriebsanleitung

Tdr-füllstandsensor
Inhaltsverzeichnis
Verfügbare Sprachen

Verfügbare Sprachen

EN
Operating instructions
SimCur for current output
SimVol for voltage output
Note: The simulation is automatically deactivated if the supply voltage is interrupted.
Simulating the level
Even if there is no liquid in the container yet, it is possible to select a filling level in the menu in
order to test the sensor configuration. When simulating a level value, all outputs on the LFP are set
according to the defined configuration. The function should not be selected until a configuration is
complete.
Configuration
Set parameter to the desired filling level as a % in the SimLev menu
Note
Simulation of the level refers to the probe length or to the liquid level in tank (probe length +
offset) if an offset was configured (please see „5.3 Testing the configuration")
The simulation is only active when there are no error messages. The simulation is automatically
deactivated if the supply voltage is interrupted.
Parameter selection
SimOff: Off
Filling level 0 %
Filling level 25 %
Filling level 50 %
Filling level 75 %
Filling level 100 %

5.4 Configuring the probe length

Log in to expert mode (Please see „5.1 Expert Mode")
Access the EXPRT-CONFIG-Length menu using the arrow and Set pushbuttons
Enter the probe length in the Length menu. Please note the defined probe length in Chapter 9.8
"Dimensional drawings"
Note:
HiSpd: max. Length = 2,005 mm, response time < 400 ms
HiAcc: max. Length = 6,005 mm, response time < 2800 ms

5.5 Programming static interference signals

Static interference signals in the tank generated by tubes, beams, couplings, or a cleaning ball can
be taught-in. The probe length provides the value for the teach-in depth.
Log in to expert mode (Please see „5.1 Expert Mode")
Access the EXPRT-Config-CalRng menu using the arrow and Set pushbuttons.
Set the value range between 95 and 6,005 mm
The following information applies:
The value starts from the LFP process connection
The value should cover all interference signals
Maximum value = probe length – 100 mm
AutCal function must be launched afterwards (see Chapter 2 "Commissioning")
76
8014485/YE12/2014-08-20 © SICK AG • Level sensors • Germany • Subject to change without notice