Step 2 — Remove the Bulb Cover
Step 3 — Locate the Lamp Holder
Step 4 — Remove Lamp Holder
Unscrew and remove the two metal lamp fasteners
from the lamp holder. Slide the lamp holder out of
the unit and place it on flat surface.
Step 5 — Lamp in Holder
Pull the old lamp out of the lamp holder.
Carefully remove the new replacement lamp from
its packaging and place in holder. Do not use
unprotected hands. Read Caution below.
CAUTION: Be sure NOT to touch the new lamp
with unprotected hands. Use tissue, soft cloth, or
non-powdered surgical glove to handle the lamp.
Minute amounts of skin oils or other impurities
with cause the lamp to fail prematurely.
FWT
31
AcuCam Concept IV
Operation Manual and Installation Guide