Table 206 - Recommended separation distances between portable and mobile RF communications
equipment and the
AcuCam® Concept® IV
portable and mobile RF communications equipment and the
The AcuCam® Concept® IV
which radiated RF disturbances are controlled. The customer or the user of the Concept IV
prevent electromagnetic interference by maintaining a minimum distance between portable and mobile RF
communications equipment (transmitters) and the Concept IV
the maximum output power of the communications equipment.
Rated maximum output
power
of transmitter
W
0,01
0,1
1
10
100
For transmitters rated at a maximum output power not listed above, the recommended separation distance d in meters
(m) can be estimated using the equation applicable to the frequency of the transmitter, where P is the maximum output
power rating of the transmitter in watts (W) according to the transmitter manufacturer.
NOTE 1 At 80 MHz and 800 MHz, the separation distance for the higher frequency range applies.
NOTE 2 These guidelines may not apply in all situations. Electromagnetic propagation is affected by absorption and
reflection from structures, objects and people.
FWT
intraoral camera system
Recommended separation distances between
AcuCam® Concept® IV
FWT
intraoral camera system is intended for use in an electromagnetic environment in
Separation distance according to frequency of transmitter
150 kHz to 80 MHz
0,12
0,38
1,2
3,8
12
FWT
intraoral camera system
FWT
camera system as recommended below, according to
m
80 MHz to 800 MHz
0,12
0,38
1,2
3,8
12
FWT
camera system can help
800 MHz to 2,5 GHz
0,23
0,78
2,3
7,3
23