Repairs and maintenance
You have purchased a product that is durable and easy to service. To guarantee faultless func-
tioning and optimal water quality, the unit must be maintained at regular intervals (at least once
per year) by an authorised service technician.
If the product malfunctions during the warranty period, contact your contract partner or the in-
stallation company, and quote the unit type and serial number (see technical specifications or the
type plate on the unit).
i NOTE!
Before performing work on electrical components or opening the housing, it is mandatory that
the mains plug be pulled and both the water infeed and the permeate line be closed off in order
to guarantee that the unit is voltage-free.
Whenever maintenance is performed, the connection lines and the device must be checked for
damage.
8.1 Wearing parts
Wearing parts must be replaced within the prescribed maintenance intervals.
Maintenance work
General visual inspection
Leak-tightness inspection
Cleaning with a damp cloth
Conductivity (with external measuring
device)
Replacement of the external
pre-filter insert (particle filter [option-
ally available])
LED flashing mode
ON
OFF
x8
ON
OFF
The status LED flashes
8 times in a row.
ON
OFF
x10
ON
OFF
The status LED flashes
10 times in a row.
Responsibility
Customer
Customer
Customer
Customer/service At least once per year
Customer/service Depending on the pre-filter used
Cause
Problems with the
membrane conducti-
vity, it does not fall be-
low the set threshold.
Conductivity limit of
the demineralisation fil-
ter exceeded.
bestdemin Plus demin-
eralisation filter cart-
ridge is likely to need
changing soon.
Interval
Weekly
Weekly
As necessary
Measure
Turn off the unit to re-
place membrane. Tab
the red button to reset.
Replace bestdemin Plus
and press the red sym-
bol.