20
The top service cover can be removed to access the two integrated filter heads with female
threads for holding the two filter cartridges:
12
Filter head for the bestaqua 14 Membrane with bypass setting, which must remain at position 0
13
Filter head for bestdemin Plus without bypass setting
1.3 Explanation of warnings
H DANGER!
Indicates a potentially dangerous situation that can lead to health impairments if not avoided.
I WARNING!
Indicates a potentially dangerous situation that can lead to health impairments if not avoided.
I CAUTION!
Indicates a potentially dangerous situation that can lead to property damage if not avoided.
i NOTE!
Highlights recommendations and information for efficient, trouble-free operation.
1.4 Symbols on the type plate
Pressure
Temperature
Date of installation and replacement of the filter cartridge
2
Technical data
2.1 Dimensions, connections and operating conditions of the ROC Clinic One
Electrical connection / fuse
Internal fuse
Mains supply voltage fluctuation
Overvoltage category
Electrical power consumption (operation/standby)
Plug standard (grounded PE mains plug)
Protection class
Water connection inlet
Water connection outlet (demineralised water)
Water connection for concentrate
Water connection for external tank
Dimensions: Width, depth, height (W×D×H)
Weight, dry
Weight, wet
Use
Max. operating altitude
Typical capacity
Typical flow rate
220-240 V / 50-60 Hz/ 10 A
max. ± 10 % of the nominal voltage
200 W / < 3 W
IEC-320 socket
3/4" external thread
3/8" external thread
8 mm connector
8 mm connector
277 mm x 297 mm x 505 mm
For indoor use only
T1.25AL250V
II
IP21
19.0 kg
22.6 kg
2000 m
3