App screen
BWT 14ROC CLINIC TP
leakage
0.1 l
20
[mmH2O]
160
TRANSPORT PUMP
II
BWT 14ROC CLINIC TP
no water
0.1 l
20
[mmH2O]
160
TRANSPORT PUMP
II
BWT 14ROC CLINIC TP
RO
pump
cooling
0.1 l
20
[mmH2O]
160
TRANSPORT PUMP
II
BWT 14ROC CLINIC TP
replace
best
demin
0.1 l
20
[mmH2O]
160
TRANSPORT PUMP
II
LED flashing mode
ON
OFF
ON
OFF
Status LED lights up in
red and a permanent
warning signal sounds
ON
OFF
x2
x2
x2
ON
OFF
The status LED re-
peatedly flashes twice
in succession
ON
OFF
x1
x1
x1
x1
x1
ON
OFF
The status LED flashes.
A warning tone sounds
three times in succes-
sion. The status LED
lights up in red.
Cause
Water leakage inside
the device or condens-
ation
No flow or insufficient
flow
Motor overheating pro-
tection has tripped.
The pump may be
broken.
Planmeca Patrol 5 has
not been replaced, it is
likely to need repla-
cing. The conductivity
no longer falls below
the limit value.
Measure
Disconnect the device
from the power and
water supply.
Contact your service
technician.
Check the water pipe
parameters.
Check the inlet valve.
Check that the pre-fil-
ter is not blocked.
The device will restart
automatically.
The device starts by it-
self when the motor
has cooled down.
The device no longer
starts by itself if the
pump is broken.
Replace Planmeca
Patrol 5 and press the
red symbol.