Battery is not rechargeable.
– To replace the battery, insert fingernail between upper and lower caps and run around join to release.
– Remove battery. Insert new battery under retaining clip, ensuring writing on battery is facing outwards.
– Replace the caps by pressing upper and lower caps together.
Battery type: CR1616. Replacement batteries are available from customer support upon request.
CLEANING
The i-Digits wristband, digits, cover and electrode surface can be cleaned with a soft damp cloth and mild soap.
Isopropyl alcohol can also be used to help with disinfection.
Clean the electrode surface regularly.
Do not use any strong chemicals.
NOTE: Do not submerge the i-Digits, wristband, cover, electrode surface or charger in water.
MAINTENANCE
– Always turn off the i-Digits when not in use.
– Charge the battery after use.
– Ensure the i-Digits is serviced every 12 months.
– For all service or maintenance, contact your prosthetist.
TROUBLESHOOTING
Contact your clinician should any problems occur affecting the use of the i-Digits.
TECHNICAL SPECIFICATIONS
i-Digits
Voltage
Max. current
Battery capacity
Max device load
Finger carry load (static limit)
Time from full open to full close
Expected service life
Power Supply
Manufacturer
Model No.
Input
Output
7.4 V (nominal)
5 A
Rechargeable lithium polymer 7.4 V (nominal); 800 mAh
capacity
20 kg/ 44lbs
5 kg/ 11lbs
0.8 seconds
5 years
XP Power
VEP36US12
100-240Vac, 50-60Hz, 0.9A Max
12VDC, 3A
8