26
Change mains input fuses
Switch off the mains switch and separate the device from the mains!
A
B
27
Returning Devices
Before returning the device, a transport securing device has to be installed.
If the device or its accessories are returned to Andreas Hettich GmbH & Co. KG, in order to provide protection for
people, the environment and materials, it has to be decontaminated and cleaned before being shipped.
We reserve the right to refuse contaminated devices or accessories.
Costs incurred for cleaning and disinfection are to be charged to the customer.
We ask for your understanding in this matter.
AB1203DEENFRIT
The fuse holder (A) with the mains input fuses is located next to the mains switch.
•
Remove the connecting cable from the machine plug socket.
•
Press the snap-fit (B) against the fuse holder (A) and remove.
•
Exchange defective mains input fuses.
Only use fuses with the rating defined for the type. See the following
table.
•
Reinsert the fuse holder until the snap-fit clicks shut.
•
Reconnect the device to the mains supply.
Model
Type
MIKRO 185
1203
MIKRO 185
1203-01
Rev. 08 / 12.2022
Fuse
Order no.
T 3,15 AH/250V
E997
T 6,3 AH/250V
2266
EN
81/154