Herunterladen Inhalt Inhalt Diese Seite drucken

Faults; Change Mains Input Fuse; Acceptance Of The Centrifuges For Repair - Hettich EBA 21 Bedienungsanleitung

Vorschau ausblenden Andere Handbücher für EBA 21:
Inhaltsverzeichnis

Werbung

Verfügbare Sprachen

Verfügbare Sprachen

EN
23

Faults

If the fault cannot be rectified according to the faults table, Hettich customer services must be informed.
Please state the type of centrifuge and the factory serial number. Both values are visible on the centrifuge type plate.
MAINS RESET:
- Mains switch OFF for longer than 10 secs.
- Mains switch ON.
Message / fault
No display
TACHO - ERROR
IMBALANCE
CONTROL - ERROR
06 - 09
N > MAX
N < MIN
ROTORCODE
MAINS INTERRUPT
VERSIONS-ERROR
CONTROL-ERROR
21 - 27
SER I/O - ERROR
30 - 38
FU / CCI - ERROR
60 - 83
N > ROTOR MAX
24

Change mains input fuse

Switch off the mains switch and separate the centrifuge from the mains!
A
B
25

Acceptance of the centrifuges for repair

If the centrifuge is returned to the manufacturer for repair, it must be decontaminated and cleaned to protect persons,
environment and material.
We reserve the right to accept contaminated centrifuges.
Costs incurred for cleaning and disinfection are to be charged to the customer.
We ask for your understanding in this matter.
You will find further information on our homepage: http://www.HettichLab.com
26/55
Cause
---
No voltage
Mains input fuses defective.
01
Faulty speedometer.
02
No rotor installed.
Defective motor, frequency converter or
drive.
---
Imbalance about motor axis through weight
differential in rotor assembly.
04,
Error in lid locking or lid closure.
05
Rotation too fast
13
Rotation too slow
10
Incorrect rotor coding
---
Power failure, centrifuging not properly
completed
12
Mismatch between electronic components
Error / defective control unit
Error / defective interface
Error / defective motor control
---
Speed in the selected program greater than
the maximum speed of the rotor.
The fuse holder (A) with the mains input fuses is located next to the mains switch.
Remove the connecting cable from the machine plug socket.
Press the snap-fit (B) against the fuse holder (A) and remove.
Exchange defective mains input fuses.
Only use fuses with the rating defined for the type. See the following
table.
Reinsert the fuse holder until the snap-fit clicks shut.
Reconnect the centrifuge to the mains supply.
Model
Type
EBA 21
1004, 1004-30
EBA 21
1004-01, 1004-31
Remedy
− Check distribution voltage.
− Check mains input fuses.
− Mains switch ON.
− Open lid.
− Turn rotor manually.
− MAINS-RESET when power is
switched on, rotor should turn.
− Open lid.
− Correct imbalance.
− Open lid.
− MAINS-RESET
− Open lid.
− Push
button.
START
− Open lid.
− MAINS-RESET
− Check the set speed.
Reduce the set speed
Fuse
Order no.
T 3,15 AH/250V
E997
T 6,3 AH/250V
2266

Werbung

Inhaltsverzeichnis
loading

Inhaltsverzeichnis