Change mains input fuses

Switch off the mains switch and separate the device from the mains!
A
B
19

Maintenance and servicing

The device can be contaminated.
Pull the mains plug before cleaning.
Before any other cleaning or decontamination process other than that recommended by the manufacturer is
applied, the user has to check with the manufacturer that the planned process does not damage the device.

Centrifuges, rotors and accessories must not be cleaned in rinsing machines.

They may only be cleaned by hand and disinfected with liquids.

The water temperature must be between 20 – 25°C.

Only detergents/disinfectants may be used which:
 have a pH between 5 - 8
 do not contain caustic alkalis, peroxides, chlorine compounds, acids and alkaline solutions

In order to prevent appearances of corrosion through cleaning agents or disinfectants, the application guide from
the manufacturer of the cleaning agent or disinfectant are absolutely to be heeded.
20

Disposal

Before disposal, the device must be decontaminated and cleaned to protect people, the environment and property.
When you are disposing of the device, the respective statutory rules must be observed.
Pursuant to guideline 2002/96/EC (WEEE), all devices supplied after August 13, 2005 may not be disposed as part of
domestic waste. The device belongs to group 8 (medical devices) and is categorized in the business-to-business
field.
The icon of the crossed-out trash can shows that the device may not be disposed as part of domestic
waste.
The waste disposal guidelines of the individual EC countries might vary. If necessary, contact your
supplier.
32/59
The fuse holder (A) with the mains input fuses is located next to the mains switch.

Remove the connecting cable from the machine plug socket.

Press the snap-fit (B) against the fuse holder (A) and remove.

Exchange defective mains input fuses.
Only use fuses with the rating defined for the type. See the following
table.

Reinsert the fuse holder until the snap-fit clicks shut.

Reconnect the device to the mains supply.
Model
Type
EBA 20
2002
EBA20
2002-01
Fuse
Order no.
T 1,6 AH/250V
E891
T 3,15 AH/250V
E997