10.3.4 Problems with the
Centrifuge?

10.4 Error Mode

10.4.1 Error Codes

– Switch off the mains power switch and disconnect the power cord from the socket.
– Pull the centrifuge a bit over the edge of the table; if necessary, jack it up
(e.g. on wooden logs). Ensure that the centre of gravity of the centrifuge is still
located above the tabletop.
– Insert the wrench for the emergency release into the hole (see fig. item 1) and
attach it to the shaft of the lid lock motor.
– Turn the shaft carefully clockwise with the aid of the wrench until the lid opens.
As a general rule, the lock can be opened more easily with the emergency lid
release system if the load on it is relieved by exerting slight pressure on the lid.
– Remove the wrench and pull the centrifuge back on the table.
Please contact your supplier for support or in the event of malfunctions
and for supply of spare parts.
In the error mode "ERR" is displayed. The error codes are displayed on
the time display.
Error no.
Kind of error
1
Tacho signal disturbed.
4
Lid does not open after pressing the lid key. Power off | on *
2, 3, 5-18
Internal fault.
19
Power supply failure during run.
* If the display changes between "ERR" and "SAFE", wait until "OFF" is displayed.
Then switch "Power on | off".
Should it not be possible to repair the failure, please contact Service!
Actions
Power off | on *
Emergency lid release.
Power off | on *
Open lid after standstill
of the rotor
Confirm error message
by pressing the lid-key
Appendix
29