5.2.8 Short Run

5.2.9 Soft Stop | Soft Start
Function
22
Operating Elements
– Press and hold the "Start | Stop" key for the whole short-run. The centrifuge accel-
erates to its maximum speed, the time run is counted in seconds. As soon as the
"Start | Stop" key is released, time is no longer counted, the centrifuge decelerates
to standstill. This mode is indicated by display of "short run".
During short run the time is counted up in seconds.
After termination of the short run the elapsed run time is displayed until the centri-
fuge lid opens automatically.
The following options are available (see 5.2.2 Speed | Relative Centrifugal Force | Time)
– Soft = soft acceleration and soft deceleration
– Stop = standard acceleration and deceleration
– Soft stop = soft deceleration
The "soft stop" or "soft" function results in a longer deceleration and | or acceleration
time. Active "soft stop", "stop" or "soft" functions are displayed.
Activation | deactivation are possible as desired.