5.2.3 Alteration of Parameters
during a Centrifuge Run

5.2.4 Time

5.2.5 Changing the Runtime
during Centrifugation

5.2.6 Standard Run Mode

5.2.7 Continuous Run Mode

– To change a parameter during the run, acitvate the edit mode using the "Select"
key. One area of the display is flashing.
– Select the desired parameter using the "Select" key and change the parameter
using the arrow keys. The parameters are entered immediately.
In the lower line the set total run time is displayed respectively the remaining run
time in the run mode (see 5.2.2 Speed | Relative Centrifugal Force | Time).
In the short run mode the total run time is counted up in seconds.
The preselected runtime can be changed during centrifugation.
– Press the "Select" key repeatedly until the time unit flashes on the display.
– Change the desired runtime by pressing the arrow keys.
The parameters will be accepted immediately.
If the centrifugation time is changed during the run, the centrifuge will run for
the entire new time and will disregard the previous runtime that has already
elapsed.
The maximum total run time is 99 minutes 59 seconds which is displayed before the
centrifuge has been started.
During centrifugation this time is counted down. The centrifuge run can be
terminated early by pressing the "Start | Stop" key. The remaining run time remains
on the display until the centrifuge lid opens automatically (see 5.2.2 Speed | Relative
Centrifugal Force | Time).
In the continuous run mode the run time is not limited and has to be terminated
manually. The centrifuge accelerates up to the set speed after pressing the
"Start | Stop" key.
The display of "--:--" indicates the continuous run mode. Unlike the standard run
mode the time is counted up in the continuous run mode.
After 99 minutes 59 seconds the run time is no longer displayed but the run contin-
ues. The display now shows "--:--".
The continuous run is terminated by pressing the "Start | Stop" key and the centrifuge
decelerates up to standstill of the rotor.
The continuous run mode can be changed to a normal run by selecting a set value
(see 5.2.2 Speed | Relative Centrifugal Force | Time).