APPENDIX A: ELECTROMAGNETIC COMPATIBILITY – GUIDANCE AND MANUFACTURER'S
DECLARATIONS
TABLE 4: Recommended Separation Distances Between Portable and Mobile RF Communications Equipment and the ICP Monitor .
Deviations from emission and immunity environment for this device may affect its life expectancy.
The ICP Monitor is intended for use in an electromagnetic environment in which radiated RF disturbances are controlled. The
customer or the user of the ICP Monitor can help prevent electromagnetic interference by maintaining a minimum distance
between portable and mobile RF communications equipment (transmitters) and the ICP Monitor as recommended below,
according to the maximum output power of the communications equipment.
Rated maximum output power
of transmitter W
0.01
0.1
1
10
100
Separation distance according to frequency of transmitter m
150 kHz to 80 MHz
D= (3.5/V1)(√ P)
.012
.037
1.17
3.69
11.67
80 MHz to 800 MHz
D= (3.5/E1)(√ P)
.012
.037
1.17
3.69
11.67
36
800 MHz to 2.5 GHz
D= (7/E1)(√ P)
0.23
.074
2.33
7.38
23.33