Connect the CS to your system as follows:

6.3.3
1.
First connect the return-flow line (not supplied) to the OUTLET port of the CS.
Thread: G1/4 ISO 228, recommended internal diameter of line: ≥ 4mm
2.
Now connect the other end of the return-flow to the system tank, for example.
3.
Check the pressure at the measurement point; it has to be within the specified
limits. The pressure in the measurement cell may not exceed 100 bar.
4.
Now connect the measurement line (not supplied) to the INLET port of the CS.
Thread: G1/4 ISO 228, recommended internal diameter of line (to prevent particle
sedimentation): ≤ 4mm
5.
Now connect the other end of the measurement line to the measurement point.
6.
The hydraulic installation of the CS is now complete.
HYDAC Filtertechnik GmbH
BeWa CS1000 3247149-02j en.doc
All manuals and user guides at all-guides.com
If there are particles larger than 400 microns in the hydraulic
system, or this is to be expected, a strainer has to be installed
upstream of the CS1000.
Oil starts to flow through the CS as soon as it has been connected
to the pressure fitting. This is why it is necessary that connection be
done in the sequence specified above.
en
Installation and Operation
Page 92
2006-01-04