Herunterladen Inhalt Inhalt Diese Seite drucken

Fluid Temperature (Temp); I/Temperature Analog Output Characteristic - HYDAC International ContaminationSensor CS 1000 Serie Betriebs- Und Wartungsanleitung

Vorschau ausblenden Andere Handbücher für ContaminationSensor CS 1000 Serie:
Inhaltsverzeichnis

Werbung

Verfügbare Sprachen

Verfügbare Sprachen

ContaminationSensor CS
Fluid Temperature (
9.1.5
The temperature signal corresponds to the fluid temperature.
9.1.5.1

I/Temperature analog output characteristic

Current I
I< 4.0 mA
I = 4.0 mA
...
I = 7.2 mA
I = 8.48 mA
I = 9.76 mA
I = 11.04 mA
I = 12.32 mA
I = 13.60 mA
I = 14.88 mA
I = 16.16 mA
I = 17.44 mA
I = 18.72 mA
I = 20.0 mA
Range 4.0 to 20.0 mA corresponds to -25°C to +100°C or -13°F to 212°C
(resolution 1°C or 1°F)
• The current can be calculated for a given temperature in °C as follows:
I = 4.0 mA + (temp+25) * (20 mA - 4 mA) / 125
• The temperature in °C can be calculated for a given current as follows:
Temp = ((I-4.0 mA)*(125/16 mA)) - 25
• The current can be calculated for a given temperature in °F as follows:
I = 4.0 mA + (temp+13) * (20 mA - 4 mA) / 225
• The temperature in °F can be calculated for a given current as follows:
Temp = ((I-4.0 mA)*(225/16 mA)) – 13
HYDAC Filtertechnik GmbH
BeWa CS1000 3247149-02j en.doc
All manuals and user guides at all-guides.com
TEMP
)
Description
Cable break
-25 °C
...
0 °C
+10 °C
+20 °C
+30 °C
+40 °C
+50 °C
+60 °C
+70 °C
+80 °C
+90 °C
+100 °C
en