Herunterladen Inhalt Inhalt Diese Seite drucken

I/Iso Analog Output Characteristic - HYDAC International ContaminationSensor CS 1000 Serie Betriebs- Und Wartungsanleitung

Vorschau ausblenden Andere Handbücher für ContaminationSensor CS 1000 Serie:
Inhaltsverzeichnis

Werbung

Verfügbare Sprachen

Verfügbare Sprachen

ContaminationSensor CS
9.1.3.2

I/ISO analog output characteristic

The current (4-20 mA) corresponds to an ISO class or an error, which is described in the
following table and diagram:
Current I
I < 4.0 mA
4.0 mA < I < 4.1 mA
4.1 mA < I < 4.3 mA
4.3 mA < I < 4.5 mA
4.5 mA < I < 4.8 mA
I = 4.80 mA
I = 5.37 mA
I = 5.94 mA
I = 6.51 mA
I = 7.08 mA
I = 7.65 mA
I = 8.22 mA
I = 8.79 mA
I = 9.36 mA
I = 9.93 mA
I = 10.50 mA
I = 11.07 mA
I = 11.64 mA
I = 12.21 mA
I = 12.77 mA
I = 13.34 mA
I = 13.91 mA
I = 14.48 mA
I = 15.05 mA
I = 15.62 mA
I = 16.19 mA
I = 16.76 mA
I = 17.33 mA
I = 17.90 mA
I = 18.47 mA
I = 19.04 mA
HYDAC Filtertechnik GmbH
BeWa CS1000 3247149-02j en.doc
All manuals and user guides at all-guides.com
Description
Cable break
Device error, device not ready
Not defined
Flow error (flow too low)
Not defined
ISO 0
ISO 1
ISO 2
ISO 3
ISO 4
ISO 5
ISO 6
ISO 7
ISO 8
ISO 9
ISO 10
ISO 11
ISO 12
ISO 13
ISO 14
ISO 15
ISO 16
ISO 17
ISO 18
ISO 19
ISO 20
ISO 21
ISO 22
ISO 23
ISO 24
ISO 25
en
Installation and Operation
Page 108
2006-01-04

Werbung

Inhaltsverzeichnis
loading

Diese Anleitung auch für:

Cocos 1000

Inhaltsverzeichnis