Herunterladen Inhalt Inhalt Diese Seite drucken

Drilling The Fastening Holes Of The Ceiling Tube; Assembly Of The Ceiling Holder - Derungs Dmed Triango 100 C Duo Montage- Und Gebrauchsanweisungen

Untersuchungsleuchte
Vorschau ausblenden Andere Handbücher für Dmed Triango 100 C Duo:
Inhaltsverzeichnis

Werbung

Verfügbare Sprachen
  • DE

Verfügbare Sprachen

  • DEUTSCH, seite 4
ENG
3.3
Drilling the fastening holes of the
ceiling tube
When determining the location of the holes, ensure that
the bore holes are aligned with the existing holes at the
lower end of the pipe.
Remove the fastening screw «C».
Insert the ceiling tube «A» into the ceiling bracket «B»
and drill the ceiling holder's existing hole to 9 mm
in diameter. Drill the opposite hole separately.
Important: When positioning the ceiling tube, ensure
that the new bore holes are aligned with the existing
holes at the lower end of the pipe.
3.4

Assembly of the ceiling holder

Fastening hardware does not come with delivery.
WARNING
Assembly by qualified personnel
Assembly must be done by qualified personnel only.
Lack of appropriate knowledge may pose a death
hazard.
Two people are needed for assembly.
20
A
C
B
DANGER
Death hazard from a falling luminaire.
The ceiling mount must only be mounted on ceilings
with a concrete stability class of B25 (C20/25)
or higher.
There should be no contact with reinforcement parts
of the solid ceiling during mounting. When in doubt,
a licensed technician must approve installation on the
specific installation surface. The load bearing capacity
of the ceiling structure must be planned, checked and
confirmed by a structural engineer beforehand.
The holes must be drilled by a professional
in compliance with the drilling tolerances permitted by
the manufacturer of the fastening anchor. If a drilling
error occurs, such as drilling into a reinforcement bar,
a structural engineer must be consulted.
The luminaire should be assembled in such a way
as to avoid permanent strain on the height stops
during operation.
If there is plastering or cladding over the concrete,
make sure the fixing anchor is properly attached right
into the concrete.
The screws must be tightened carefully using a torque
wrench according to the specifications of the fastening
anchor manufacturer.
DANGER
Danger to life through an electric shock.
All poles of the power cable must be disconnected
from the power grid from an external lockable switch.
D
Before drawing the bore holes, observe alignment of the
ceiling plate.
Alignment of the holes «D» according to the alignment
of the bearing head at the end of the pipe.
D

Werbung

Inhaltsverzeichnis
loading

Inhaltsverzeichnis