CAREFUL
Material damage due to incorrect cleaning
For cleaning, use only products that do not impair the
function of the luminaire.
For cleaning, do not use solvent-based, chlorine-
based or abrasive detergents, because they can crack
plastic parts and cause other damage.
The cleaning agents must be approved for use on
plastics such as PC, PMMA, PA and ABS.
Concentrated disinfectant can damage the luminaire.
For concentration and application times, check the
information provided with the product used.
The wrong cloth may cause scratches.
RECOMMENDED DISINFECTANTS
Lysoformin
®
Dismozon
®
Hexaquart
plus
®
Sagrotan
quick disinfectant cleaner
®
CAREFUL
Dirt reduces the light strength.
Keep cover clear through regular cleaning.
Only wipe cleaning allowed.
Clean the PA clear cover with a suitable non-abrasive
cleaning cloth and a suitable cleansing agent.
CAREFUL
Wipe cleaning is only allowed in a horizontal position
24
CAREFUL
To minimize the risk of disease transmission, applicable
health and safety regulations and the requirements of the
national hygiene and disinfection authorities must be
observed in addition to these instructions.
6.1

Sterilizing the hand grip

Sterilization must be done according to ISO 17665-1
(Sterilization of health care products in moist heat).
CAREFUL
Damage to the hand grip
Do not sterilize with hot air
Package the hand grip in a sterile bag before
sterilization.
The hand grip is designed exclusively for damp
sterilization with 3 times fractioned pre-vacuum and
saturated steam with the following parameters:
Temperature
Overpressure
Dwell time
Vacuum drying
After sterilization, check the hand grip for mechanical
integrity.
Do not use damaged hand grips.
7.

SAFETY INSPECTIONS

DANGER
Danger to life through an electric shock.
Unplug from the grid and turn switch to off position.
The power supply cable must be checked at least
once a year for damage.
CAREFUL
Maintenance and repairs can be performed only by
qualified electricians.
The corresponding user profile is in Chapter 2 Safety
instructions.
ANNUALLY:
Check the power supply cable for damage and
replace if necessary
Check for deformations and cracks in the plastic parts.
Check the load bearing system for distortion or
damage.
Check for loose parts.
134°C
2.0 bar
6 min
20 min