Herunterladen Inhalt Inhalt Diese Seite drucken

Sensor Cable Replacement; Change Of Reflector - Henkel LOCTITE UVALOC 1000 Bedienungsanleitung

Inhaltsverzeichnis

Werbung

Verfügbare Sprachen

Verfügbare Sprachen

8
Troubleshooting
• Insert the sensor cable plug into the
relevant socket on the PCB.
• Re-assemble top cover and tighten the 4
screws.
CAUTION!
Be careful not to disturb the ground cable!
8.6

Sensor Cable Replacement

• To disassemble lamp housing from the cure chamber, remove the 4 screws.
• Lift lamp housing off.
• Disconnect plug connection.
• Disassemble base plate - remove the 10 screws.
• Disconnect sensor cable from the connector on the sensor PCB and connect plug of
new cable.
• Pass both sensor cables through one of the rear ventilation slots into the lamp housing.
• Re-assemble base plate.
• Re-connect plug connection.
• Re-assemble lamp housing to the cure chamber and screw down.
• Remove the 4 screws which hold down the red top cover.
• Lift the top cover off.
CAUTION!
Be careful not to disturb the ground cable!
• Disconnect sensor cable from the plug on the PCB and connect plug of new cable.
• Re-assemble top cover and tighten the 4 screws.
8.7

Change of Reflector

• Unlock and open the drawer.
44