Connection J21 Accessories and Options
Interconnections
The controller is provided with a power connector plug J21. Ensure
that the main circuit breaker is off, then remove the plug and wire
the pins (maximum wire size 18 AWG) as indicated below, to obtain
the desired capability as shown in the following figure.
Figure 14
Pin 1a - Ground
Pin 1b-3a, 3b-3a, 2b-2a -Output of 120 Vac (independent of line
voltage), 1 A total for vent device and forepump relay coil.
Voltage is present after START push button is pressed and remains
present until an overload condition occurs or power is removed from
the turbo-pump.
Pin 4b-4a -Output mains voltage 3 A max. for heater band. Voltage is
present when the turbopump is running and the heater band enable
is selected YES.
Turbo-V 6000 Controller User Manual / 87-900-855-01

Connection J21 Accessories and Options Interconnections

Power socket output signals
Technical Information