Error Messages

For a certain type of failure, the controller will self-diagnose the
error and the messages described in the following table are
displayed.
Tab. 1
MESSAGE
CHECK
CONNECTION
TO PUMP
PUMP
WAITlNG
INTERLOCK
FAULT:
OVERTIME
FAULT: PUMP
OVERHEATED
FAULT:
CONTROLLER
OVERHEATED
FAULT:
OVERLOAD
Turbo-V 6000 Controller User Manual / 87-900-855-01 (F)
DESCRIPTION
Wrong connection between the
pump and the controller (P31).
The interlock signal of P1 connector
is activated by an interruption of
the link between pin 3 and 8 of J1
connector, or because the external
interlock signal is open.
The pump did not reach the
expected speed of 5500 KRPM 12
minutes after startup with SOFT
START deselected.
The temperature has exceeded one
of the following threshold values:
- 65 °C for the bearings
- 90 °C for motor.
- 60 °C for the water.
The controller transformer
temperature exceeds 90 °C or the
temperature on the outgoing
Mosfets radiator exceeds
60 °C.
In normal operation (after the
starting phase) the current drawn
by the pump is higher than
programmed (25 A)
Instructions for Use

Error Messages

REPAIR ACTION
Check connection between
controller and pump.
Press the START push-button
twice to start the pump.
Reset the short circuit
between pin 3 and pin 8 of J1
connector, or close the
external interlock signal.
Make sure that the system
does not leak. Press the
START button twice to restart
the pump.
Wait until the temperature
decrease below threshold
value.
Press the START push-button
twice
to start the pump.
Wait until the temperature
decrease below threshold
value.
Press the START push-button
twice
to start the pump.
Check that the pump rotor is
free to rotate.
Press the START push-button
twice to start the pump.
203/274
16

Werbung

loading