Instructions for Use
Error Messages
MESSAGE
FAULT: SHORT
CIRCUIT
SYSTEM
OVERRIDE
OVERVOLTAGE
OIL LEVEL AT
MIN
204/274
DESCRIPTION
After the starting phase the output
connection is shorted (output
current higher than 60 A).
The pump is stopped by an
emergency stop signal provided via
a remote contact.
Controller power supply circuitry is
faulty, or the Controller received a
spike.
The oil level sensor detects a level
less than the minimum allowed.
Turbo-V 6000 Controller User Manual / 87-900-855-01 (F)
REPAIR ACTION
Check connections and
shortages between pump and
controller. Press the START
push-button twice
to start the pump.
Check the insulation between
motor and pump.
Remove the controller power
cable and check the
emergency condition. Then
reconnect the power cable
and
press the START push-button
twice
to start the pump.
Press the START push-button
twice to start the pump.
Should the message still be
present, call the Agilent
service
Carry out pump maintenance
as provided by the manual.