Connection J7 – Accessories and Options
Interconnections
Figure 13
The output voltage of the connector J7 is 120 Vac and the maximum current is
CAUTION!
1A. Do not connect this plug to a device with a leakage current greater than
1.75 mA.
The 120 Vac, 1 A maximum (independent of line Voltage) output
Voltage is present when the main power switch is set to 1 position
and after START push-button is pressed and will remain present
until a fault condition is displayed on the front panel display or the
turbopump is stopped.
The connector J7 is for vent device, forepump connector coils, pump
cooling fan etc.
To make connections, remove the plug and wire the pins (maximum
wire size 18 AWG, 1 mm²) as indicated in the figure to obtain the
desired capability.