Diagnostic buffer

OFF
MEANING
SET
0
B
station non exist (set by master)
0
B
station not ready (slave not ready for
1
data exchange)
B
cfg fault (configuration data doesn't
2
match)
B
ext diag (slave has external diagnostic data)
3
B
not supported (slave doesn't support re-
4
quested function)
B
invalid slave response ( slave sets per-
5
manent '0')
B
prm fault (wrong parameter assignment)
6
B
master_lock (slave is parameterised by
7
another master)
1
B
prm req (slave has to be re-
0
parameterised)
B
stat diag (static diagnosis)
1
B
fixed '1'
2
B
wd_on (response monitoring active)
3
B
Freeze mode (received freeze command)
4
B
Synch mode (received synch command)
5
B
reserved
6
B
diag deactivated (slave is parameter-
7
ised by another master)
2
Reserved
3
Master Add (master address after parameter-
isation. FF= without parameter)
4
Ident number high
5
Ident number low
6
External diagnosis "Header" length indication in-
cluding header
7
Interface Status
B
Controller not responds
0
B
Controller not identified
1
B
Input data not fully coherent yet
2
B
Not used
3
B
Not used
4-7
8
B
– Start/Stop diag status
7-4
B
– Low Speed diag status
3-0
9
B
– Soft Start diag status
7-4
B
– Active Stop diag status
3-0
10
B
– Vent Valve diag status
7-4
B
– Purge diag status
3-0
11
B
– Heather Jacket diag status
7-4
B
– Stop Speed Reading diag status
3-0
12
B
– High speed setting diag status
7-4
B
– Low speed setting diag status
3-0
13
B
– Water Cooling Max Temp diag status
7-4
B
– Water Cooling Hyst diag status
3-0
14
B
– Set Point Hyst diag status
7-4
B
– Heather Jacket Setting diag status
3-0
15
B
– Flow Meter Threshold diag status
7-4
B
– Reserved diag status
3-0
16
B
– Set Point Value diag status
7-4
B
– Set Point Delay diag status
3-0
17
B
– Vent Valve Open Delay diag status
7-4
B
– Vent Open Time diag status
3-0
18
B
– Vent/Purge Auto/Man diag status
7-4
B
– Vent/Purge Remote/Profibus diag status
3-0
19
B
– Interlock Mode diag status
7-4
B
– Set Point Logic diag status
3-0
20
B
– Heather Jacket Mode diag status
7-4
B
– Load Gas Type diag status
3-0
21
B
– Purge Gas Type diag status
7-4
B
– Start/Stop/Interlock Mode diag status
3-0
22
B
– Set Point Type diag status
7-4
B
– Analog Out Type
3-0
23
B
– Reserved
7-4
B
– Reserved
3-0

RS485 CABLE INSTALLATION

The supplied serial cable must be installed when the
TV 301-P Navigator Controller have to be controlled by
means of a remote computer.
The cable is installed fixing the 9 pin D-type connector
into the P2 serial connector as shown in the following
figure.

CONNECTION CONFIGURATIONS

Connection C - ELECTRICAL

The turbopump is connected to the controller through an
6-pin connector. Pins B, C and D are the 3-phase
supply to the motor, pins A and F are connected to the
temperature sensor (NTC type, 30 KΩ resistance
at 25° C) and pin E is connected to the pump ground.
If the temperature sensor is disconnected, the pump will
not start. To prevent damage to the pump when the
temperature exceeds 60° C, the sensor automatically
cuts out the power supply.

ACCESSORIES AND SPARE PARTS

DESCRIPTION
Serial cable and Navigator
Software
Fan extension cable
50
TECHNICAL INFORMATION
PART NUMBER
969-9883
969-9949
87-901-001-01 (A)