8 In vitro performance

Figure 11. Typical hematocrit
1 Inlet hematocrit (Hct %)
2 Achievable hematocrit, mean (Hct %)
The data in Figure 11 are from in vitro testing using human blood and standard wash protocol. Clinical results may vary.

9 Transportation and storage

9.1 Transporting the device

Warning: Failure to adhere to the following loading conditions when transporting the device may lead to tipping:
• When using the bottom IV pole hooks, do not attach more than 5 L of liquid, and do not attach any bag higher than 48.3 cm (19 in)
above the device.
• When using the top IV pole hooks, do not attach more than 3 L of liquid, and do not attach any bag higher than 86.4 cm (34 in)
above the device.
• Do not attach a collection reservoir.
• Do not attach a waste bag.
Caution: Do not move the device by holding the IV pole, as this action could damage the device or cause tipping.
• Unplug the device power cord before transporting the device.
• Do not drop the device.
• Do not push, tip, or lean on the IV pole.
• If the device is tipped over or dropped, ensure that a trained service technician examines the device before use.
• Unlock the wheels if necessary.
• Only move the device while holding the handle. Rotate the handle to the push position (Figure 7).
• Use caution when transporting the device when fluids are present.
• See Chapter 21 for transit and storage environmental limits.

9.2 Transporting the device between clinical sites

• The device stores data after a password is set. Refer to institutional policies regarding the handling of protected health
information and data security before transporting the device to another clinical site.
• Before moving the device, retract the IV pole and secure it by tightening the IV pole knobs.
• Lock the wheels before lifting the device into a vehicle.
• If it is necessary to lift the device, it should be done by a minimum of 2 people. Lift the device by holding it at the device base
(Figure 5) or the power cord wrap (Figure 6). Do not lift the device by holding any other points. Use caution when lifting the
device.
• The device may be transported without its original packaging, either standing upright or lying on its back.
Warning: Use caution when transporting the device by vehicle. Do not expose the device to extremes of temperature, shock,
or vibration.

9.3 Storing the device

• Store the device according to the environmental conditions specified in Chapter 21.
• When not in use, turn off the device power and disconnect the power cord from the supply mains. Store the power cord by
wrapping it around the power cord wrap on the back of the device (Figure 6).
• Store the device in a cool, dry, and well-ventilated area.