Insert the kit in the valve and the pump.
Press
to resume.
Bowl may not be full. Wash it?
Waste bag may be full. Empty if neces-
sary. Press
to resume.
Unlock then relock the bowl cover. Press
to resume.
English
M967226A006 1B Medtronic Confidential
Response
3. Place the 3 manifold tubing lines into the tubing
guides, and ensure that each tube is fully
seated in each guide.
4. Place the single tubing line of the manifold into
the air detector, and ensure that it is fully
seated.
Note: There is a feature on the manifold cover
that pushes the tubing into the air detector
when closing the cover.
5. Rotate the manifold cover latch until the lock
indicators are aligned.
The kit is not properly installed in either the manifold
a
valve or the pump.
1. Open the manifold cover.
2. Place the manifold into the recessed cavity
until the connector is fully seated in the cavity.
3. Place the 3 manifold tubing lines into the tubing
guides, and ensure that each tube is fully
seated in each guide.
4. Place the single tubing line of the manifold into
the air detector, and ensure that it is fully
seated.
Note: There is a feature on the manifold cover
that pushes the tubing into the air detector
when closing the cover.
5. Rotate the manifold cover latch until the lock
indicators are aligned.
6. Pull the pump lever and hold it open.
7. Stretch the pump tubing around the pump
head and over the pump outlet tubing guide.
8. Release the pump lever.
This message occurs when selecting Wash, after
the prompts Air in reservoir line. Press
again or select Return, Concentrate, or Wash.
or Air in holding bag line. Bowl is not full. Press
to try again or select Return or Wash.
1. Select Yes to wash the bowl.
Warning: Choosing to wash a partially filled
bowl may result in low blood quality and low red
cell volume.
2. Selecting No stops the bowl and returns the
blood to the reservoir.
The device estimates the volume in the waste bag
and a notice appears when the bag is full.
1. Empty the waste bag.
Note: Leave 100 mL to 200 mL of fluid in the
bag to provide for proper expansion during fill-
ing and emptying.
2. Press
The bowl moved during rotation and is unlocked.
1. Unlock and remove the bowl.
2. Examine the centrifuge chamber for debris and
liquid.
3. Remove the debris or liquid if necessary.
4. Examine the bowl for damage.
5. Discard and replace the bowl if it is damaged.
6. If clots are observed in the bowl, discard and
replace the wash kit.
7. Confirm that the bowl sensor and sensor
spring are moving freely and not damaged or
dirty.
8. Reinstall the bowl.
to try
to resume.
44
Audible alert
Visual alert
signal
signal
Medium
Medium
Low
Low
Medium
Medium
Medium
Medium