Recommended separation distances between portable and mobile
RF communications equipment and the EZ-IO® Power Driver
The EZ-IO® Power Driver is intended for use in an electromagnetic environment in
which radiated RF disturbances are controlled. The customer or the user of the EZ-IO®
Power Driver can help prevent electromagnetic interference by maintaining a minimum
distance between portable and mobile RF communications equipment (transmitters)
and the EZ-IO® Power Driver recommended below, according to the maximum output
power of the communications equipment.
Separation distance according to frequency of transmitter m
Rated maximum
output power of
150 kHz to 80 MHz
transmitter
d = 3,5
P
W
V
1
.01
0.12
.1
0.38
1
1.2
10
3.8
100
12
For transmitters rated at a maximum output power not listed above, the recommended
separation distance d in meters (m) can be estimated using the equation applicable to the
frequency of the transmitter; where P is the maximum output power rating of the transmitter
in watts (W) according to the transmitter manufacturer.
NOTE 1 At 80 MHz and 800 MHz, the separation distance for the higher frequency range applies.
NOTE 2 These guidelines may not apply in all situations. Electromagnetic propagation is
affected by absorption and reflection from structures, objects, and people.
• Medical electrical equipment needs special precautions regarding EMC and needs to be
installed and put into service accordingly to the EMC information provided in this manual.
• Portable RF communications equipment should be used no closer than 30 cm (12 inches)
to any part of the EZ-IO® Power Driver including cables specified by the manufacturer
Otherwise, degradation of the performance of this equipment could result.
• The use of accessories, transducers, and cables other than those specified by the
manufacturer, may result in increased emissions or decreased immunity of the EZ-IO®
Power Driver.
• The EZ-IO® Power Driver should be observed to verify normal operation in the
configuration it will be used.
• The EZ-IO® Power Driver is designed and tested to run intermittently with a duty cycle of
10 seconds on, 1 minute off for 5 consecutive cycles. Allow 1 hour cool down time.
• Environmental Conditions of Use: -20° to + 40°C; 0 to 90% Relative Humidity
Equipment Classification
Type of protection against electric shock
Degree of protection against electric shock
Degree of protection against ingress of water
and particulate
Degree of safety or application in the presence of
a flammable anesthetic mixture
80 MHz to 800 MHz
800 MHz to 2,5 GHz
d = 3,5
P
d = 7
P
E
E
1
1
0.12
0.23
0.38
0.73
1.2
2.3
3.8
7.3
12
23
NA internal powered equipment
Type BF applied part
IP33
Equipment not suitable for use in the presence
of a flammable anesthetic mixture with air or
with oxygen or nitrous oxide

WARRANTY INFORMATION

TELEFLEX LIMITED EXPRESS WARRANTY AND DISCLAIMERS
Teleflex warrants to the original end user of the Products only ("End User") that during the applicable Warranty
Period: (a) the hardware Products will conform with Teleflex's written product specifications for such Products
in all material respects for the shorter of (i) one year after shipment to End User or (ii) the number of uses of
such hardware Product as are specified by Teleflex in its written product specifications, and (b) the disposable
Products will conform with Teleflex's written product specifications for such Products in all material respects
until the expiration date designated therefor on such disposable Products (collectively, the "Warranty
Period"), unless the Products have been subjected to physical abuse, misuse, abnormal use, use not consistent
with Teleflex's published directions and instructions for use, fraud, tampering, unusual physical stress,
negligence or accidents ("Express Warranty"). Teleflex does not guarantee that the operation of a hardware
Product will be uninterrupted or error-free. Teleflex will, in its discretion, repair, replace or refund the purchase
price to End User for Product determined by Teleflex to be non-conforming ("Remedies"), provided that End
User returns the nonconforming Product to Teleflex during the applicable Warranty Period, at End User's
expense and first gives prompt written notice to Teleflex so that Teleflex can issue a Return Material
Authorization ("RMA") number. Products sent to Teleflex for warranty replacement without a valid RMA number
displayed on the outside of the shipping container may, in Teleflex's discretion, be returned to End User at End
User's expense. All returned nonconforming Product become the property of Teleflex. To the extent permitted
by law, Teleflex may repair or replace nonconforming hardware Products (a) with new or previously used
Products or parts equivalent to new in performance and reliability, or (b) with equivalent Products to an
original Product that has been discontinued. Replacement Products (or parts thereof) are warranted for the
remainder of the Warranty Period of the Product they are replacing. THE REMEDIES DESCRIBED HEREIN SHALL
BE END USER'S SOLE AND EXCLUSIVE REMEDY FOR A FAILURE OF A PRODUCT TO CONFORM TO THE EXPRESS
WARRANTY. TO THE MAXIMUM EXTENT PERMITTED BY APPLICABLE LAW, THE EXPRESS WARRANTY IS THE SOLE
AND EXCLUSIVE WARRANTY AND GIVEN IN LIEU OF ALL OTHER WARRANTIES, WHETHER EXPRESS, IMPLIED OR
STATUTORY, INCLUDING WITHOUT LIMITATION THE IMPLIED WARRANTIES OF MERCHANTABILITY, NON-
INFRINGEMENT, FITNESS FOR A PARTICULAR PURPOSE, SATISFACTORY QUALITY OR SUITABILITY. IF THE
DISCLAIMER OF ANY IMPLIED WARRANTY IS NOT PERMITTED BY APPLICABLE LAW, SUCH EXPRESS WARRANTY
IS LIMITED TO NINETY (90) DAYS FROM THE DATE OF ORIGINAL PURCHASE. OTHER THAN THE EXPRESS
WARRANTY, THE PRODUCTS ARE PROVIDED "AS IS" AND ARE DESIGNED FOR USE SOLELY BY QUALIFIED
HEALTHCARE PERSONNEL USING REASONABLE MEDICAL DISCRETION IN MEDICALLY NECESSARY SITUATIONS.
TELEFLEX DISCLAIMS ALL LIABILITY WITH RESPECT TO THE PRODUCTS ARISING FROM ANY USE OF THE
PRODUCTS THAT IS INCONSISTENT WITH TELEFLEX'S PUBLISHED DIRECTIONS AND INSTRUCTIONS FOR USE. IN
NO EVENT SHALL TELEFLEX BE LIABLE TO END USER, ANY CUSTOMER OR ANY OTHER THIRD PARTY ("CLAIMANT")
IN ANY MANNER FOR ANY SPECIAL, NON-COMPENSATORY, CONSEQUENTIAL, INDIRECT, INCIDENTAL, STATUTORY
OR PUNITIVE DAMAGES OF ANY KIND, WHETHER ARISING UNDER CONTRACT OR TORT LAW (INCLUDING
NEGLIGENCE AND STRICT PRODUCT LIABILITY),REGARDLESS OF THE FORM OF LEGAL ACTION EVEN IF TELEFLEX
IS AWARE OF THE POSSIBILITY OF ANY SUCH DAMAGES IN ADVANCE. TELEFLEX'S TOTAL AGGREGATE LIABILITY
IN CONNECTION WITH THE PURCHASE OR USE OF THE PRODUCTS SHALL NOT EXCEED THE SUM OF THE AMOUNTS
PAID BY CLAIMANT TO TELEFLEX DURING THE TWELVE (12) MONTHS IMMEDIATELY PRECEDING THE DATE OF THE
EVENT GIVING RISE TO A CLAIM AGAINST TELEFLEX.
EZ-IO
INTRAOSSEOUS VASCULAR ACCESS

Quicklinks ausblenden:

Werbung

Inhaltsverzeichnis
loading

Inhaltsverzeichnis