Adjusting the AP:
1)
Press the "ATC" key. In the display appears the torque mode set (for example [AL 20]).
2)
Pressing the keys "▲" and"▼" will increase and decrease the torque level respectively.
3)
By pressing the "ATC SET" briefly once more the digital display switches back automatically.
If the display says [AL.....] :
press the key "▼" to get back to [AP.....].
The green light "ATC" will illuminate when the AP-Modus is activated.
The AP-Modus works only when the motor runs forward (the green light LED
"Rev/Forw" does not illuminate).

5.5 Memory

Once programmed, the preselected speed, torque and value setting will remain stored in memory, even
when the motor is turned off.
Please pay attention to the following important points for the maintenance of the device:
Clean, disinfect and sterilize the device and components after each use!
Do not use dissolving agents for cleaning
Autoclave material in transparent packaging!
Do not fill the sterilization bag more than 80%.
Autoclave material at 134° C maximum.
Sterilized material should be stored and tagged with sterilization date!

6.1 Control unit and footswitch

The control unit and the pedal do not come into contact with patients. Only clean the surface of the unit,
do not use harsh cleaners or solvents for cleaning. Use 80% ethyl alcohol or microbiologically effective
disinfectants.
The control unit should be wiped with a clean, damp cloth only. No spray cleaners
or disinfecting agents should be employed, as the control unit face is not
watertight!
6

6 Disinfection, Cleaning and sterilization

NOUVAG AG
No. 31817 29/05