Cleaning, disinfection and sterilization

The following points are important with regard to caring for the material:
● Perform cleaning, disinfection and sterilization after every treatment!
8.1

Mesh boards

● Single use mesh boards may not be reused!
● Mesh boards must be disposed off properly after use!
● Don't use mesh boards when the package was open or damaged.
● Don't use mesh boards when expiry date is timed out.
● Use only mesh boards from Nouvag AG.
8.2

Reprocessing instructions, Mesher

Reprocessing
Frequent reprocessing of the Mesher has only a limited impact. The end of the product service life is
restrictions
normally determined by wear and damage through use.
INSTRUCTIONS
At location of use
Remove soiling with a disposable cloth/paper towel.
Storage and
No special requirements. Due to the risk of drying and corrosion, reprocessing must be performed
transport
without undue delay.
Preparation for
Remove the ratchet. Turn the clamping wing-grips to open the hinged bridge. Remove the blade roll.
cleaning
Remove soiling from Mesher, blade roll and the ratchet under running water using a suitable brush.
Don't put the Mesher in an ultrasonic bath.
Automatic cleaning
Equipment: Neutral or alkaline cleaning agent in b recommended concentration.
and disinfection
1. Place Mesher, ratchet and blade roll in a fine meshed basket.
2. Set a cleaning cycle that offers sufficient cleaning and rinsing. Perform the last rinse with deionized
water.
3. Perform a 10-minutes rinse cycle at 93°C to facilitate thermal disinfection.
4. When removing the Mescher, the ratchet and the blade roll check to verify whether soiling is still
visible in gaps and grooves. If necessary, repeat the cycle or clean manually.
Manual cleaning
Equipment: Neutral or alkaline cleaning agent in the recommended concentration, suitable brush and
running water.
Procedure:
1. Rinse off and brush away surface contaminants from the Mesher, ratchet and the blade roll.
2. Use the brush to apply cleaning agent to all surfaces and gaps.
3. Rinse Mesher, ratchet and blade roll thoroughly under running water.
10
Mesher, skin expansion system
Operating instructions Mesher, Ref. 31584, V32/15