Herunterladen Inhalt Inhalt Diese Seite drucken

Starting The Grinding; During Grinding; Stopping The Grinding; Automatic Stop - struers Accutom-50 Gebrauchsanweisung

Inhaltsverzeichnis

Werbung

Verfügbare Sprachen

Verfügbare Sprachen

Starting the Grinding

During Grinding

Stopping the Grinding

Automatic Stop

Manual Stop

Continuing the Grinding

Accutom-50
Instruction Manual
Before you start grinding make sure that the grinding chamber is
clean to ensure a free flow of cooling water. The grate must not be
clogged by debris.
Position the sample correctly
Set the appropriate grinding parameters.
Make sure that the correct cup wheel is mounted
Close the cover of the machine.
Press START
.
After START
has been pressed, the display changes to the
following:
Pre-set Removal
Actual
Removal
Remaining Process Time:

Y-Feed Speed
Wheel speed
Both the pre-set and the actual removal or thickness are displayed.
Accutom-50 automatically stops the grinding process at the pre-
set stop position or after the pre-defined amount of material has
been removed.
The sample is then retracted and the cup wheel is stopped.
The grinding process can be stopped at any time during the
operation by pressing the STOP
its actual position and the process stops there.
The display changes back to what it was before grinding was
started.
If the process has been stopped manually before the specified
amount of material was removed the process can be continued by
pressing START  again. The process will start again and only the
remaining amount of material will be removed.
To start all over again press ENTER
value to the original setting.
GRINDING STARTED
:
0.100 mm
:
0.100 mm
3 min
:
10.0 mm/s
:
4000 rpm
key. The sample remains in
twice to reset the removal
31

Quicklinks ausblenden:

Werbung

Inhaltsverzeichnis
loading

Inhaltsverzeichnis