Options

Language

X-retract
Contrast
No. of Final Grinding Sweeps
F1: DEFAULT VALUE
Accutom-50
Instruction Manual
In the options menu various items can be configured.
See how to change the language in the Getting Started section of
this Instruction Manual.
After the stop position is reached, the sample is normally moved
straight back to the relative zero position. To avoid any contact with
the cut-off wheel during that movement, the sample can be retracted
from the wheel before it is repositioned by setting X-retract to ON.
After reaching the relative zero position, the sample is then moved
back into the original position.
Make sure that the Y-stop position is set correctly. If the sample is not cut
through before the sample is retracted the cut-off wheel might be damaged.
The contrast setting of the display can be adjusted.
When grinding, the number of sweeps carried out before the sample
is moved towards the cup wheel can be adjusted. To save time, very
often only a small number of sweeps is set during normal operation.
However, to achieve a better surface finish, the number of final
sweeps can be set to a maximum of 15.
Pressing F1 will reset the inverted option to the factory adjusted
default value.
IMPORTANT