Rotation

Rotation: Off
Rotation: Continuous
Rotation: Oscillating

Removal Mode

Material to be removed
Accutom-50
Instruction Manual
Sample rotation can be an advantage to get a more uniform scratch
pattern on the surface and a better planeness. On Accutom-50, three
different settings are possible.
The sample does not rotate.
The sample is rotating around its centre.
Press F3 to test sample rotation with the set parameters.
The sample is rocking/oscillating around its centre.
The angle can be adjusted from 10° to 400°.
Press F3 to test sample rotation with the set parameters.
On Accutom-50 two different ways of removing a certain amount of
material can be used.
If you know the exact position of a component underneath the
sample surface, for example 0.125 mm, this distance can be pre-set.
Move the sample as close to the cup wheel as possible, without
getting in contact with each other using the POSITION  key.
Press the POSITION
cup wheel just until there is just contact between sample and
wheel.
Press F1:RESET to set the relative X-position to zero.
After having defined the amount of material to be removed, move
the sample slightly back from the cup wheel, in the Y-direction,
, and after completing the grinding process
press START
Accutom-50 will stop precisely at the pre-defined depth.
key to move the sample slowly against the