Loading a Grinding Method
Positioning the Sample
Pre-set Amount of
Material to be Removed
Stop Grinding at Position
Starting the Process
Accutom-50
Instruction Manual
Grinding with Accutom-50
Press Esc until the Main Menu appears.
Select Grinding Methods and press Enter .
Select the method you want to use and press Enter .
Place the specimen holder with the sample in the specimen
holder head and clamp it.
If you know the exact position of a component underneath the
sample surface, for example 0.125 mm, this distance can be pre-set:
Move the sample slowly against the cup wheel just until there is
contact between sample and wheel.
Press F1:RESET to set the relative X-position to zero.
Define the amount of material to be removed, then move the
sample away from the cup wheel in Y-direction.
Grind the ceramic plate of the vacuum holder so that it is plane
parallel to the cup wheel.
After the grinding press F1:RESET to set the relative X-position
to zero.
Define the thickness of the glass plate you want to grind by
setting the Stop grinding at position: to the wanted value.
Move the holder away from the cup wheel to allow the glass
plate to be inserted.
Place the glass plate on the vacuum holder.
Reposition the glass plate close to the cup wheel.
Press START
thickness.
Press F1:RESET to set the relative X-position to zero.
Move the holder with the glass plate away from the cup wheel to
allow the sample to be inserted.
Mount the sample on the glass plate and define the thickness of
the sample by resetting the Stop grinding at position: to the new
value.
Reposition the sample close to the cup wheel.
Press START
to grind the glass down to the pre-set
to start the grinding process.